OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the NIDEK Auto Lensmeter (LM)
3.2.3 Operating procedure
1 After lens measurement with the LM, press the print button on the LM.
2 The TONOREF II receives lensmeter readings from the LM.
3 The LM data is printed and data is transmitted to the RT.
To RT
Interface cable
(optional)
To LM
Ferrite cores
(optional)
When the device communicates with the LM, set the communication parameters of each
instrument as follows. See the operator’s manual for the setting method of each device.
[Settings of TONOREF II]
83. BAUD-RATE = 9600
84. BIT LENGTH = 8
86. LM DATA PRINT = YES
[Settings of NIDEK lensmeter]
PRINTER = COM PRINT
RS-232C = NIDEK
Baud Rate = 9600
Parity = Odd
Data Bits = 8
Stop Bits = 1