BEFORE USE: Measurement Screen Description
1
Focusing indicator
Indicates the distance between the main body and the patient’s eye.
Operate the joystick until you can obtain the proper focus ( ).
Manual mode button ( )
Turns off both the auto-tracking and auto-shooting functions (manual mode).
The auto-tracking mark and auto-shooting mark become blank ( ), indicating that these functions are
turned off. Pressing the manual mode button returns to the state before the manual mode button was pressed.
See “2.4.1 Switching to manual mode” (page 37) for details on manual mode.
Ring image enlargement button ( )
Switches to the ring image full screen by pressing this button
when the thumbnail of the measurement ring is displayed after
AR measurement.
See “{ Measurement ring image display” (page 48) for details.
Mire ring
Used as an alignment reference ring.
When the auto-tracking function is on (3D or 2D), bring the mire ring close to the target so that the device
automatically starts alignment.
When the auto-tracking function is off, bring the patient’s eye into focus so that the mire ring is placed within the
target.
If the eyelid or eyelashes are on this mark, KM measurement may not be possible.
Page button ( , , )
Switches the measurement screen among Page 1, Page 2 and Page 3.
Pressing the button switches the page in the following order: Page 1 Page 2 Page 3 Page 1 ......
The displayed icons vary according to the selected page.
There is no Page 3 for NT mode.